CVE-2019-6538
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
Buzz score
Why this CVE is surfacing
This all-time snapshot uses the same composite formula as Trending across a 30-year evidence window, rather than a current rolling window.
Buzz score components · mention 11.0 · diversity 6.5 · KEV 0.0 · OTX 0.0 · PoC 0.0
Why it matters now
Mention timeline
- Total mentions
- 0
- within the 30d window
- Peak daily
- 0
- highest bucket
Evidence
Source links by recency
2 source links · newest first
The Department of Homeland Security (DHS) has issued an alert to warn of critical vulnerabilities impacting numerous Medtronic devices, which are exploitable with low skill level.
newswww.securityweek.comMar 25, 2019, 2:16 PM- Medtronic’s implantable heart defibrillators vulnerable to hackSecurity Affairs
The U.S. Department of Homeland Security Thursday issued a security advisory for multiple vulnerabilities affecting over a dozen heart defibrillators. Multiple vulnerabilities in the heart defibrillators could be exploited by attackers to remotely control the devices, potentially putting the lives of patients at risk. An implantable cardioverter-defibrillator (ICD) is a device implantable inside the human […]
newssecurityaffairs.comMar 22, 2019, 8:53 PM
Exploit code
Public exploit repository references
Public PoC repositories are third-party, potentially unsafe artifacts. Treat their code as untrusted and use it only on authorized systems in an isolated, least-privilege environment. cvebuzz does not execute the code or verify that an exploit works.
0 repository references · best confidence N/A · max 0 stars
Related records
Similar CVEs
6 related CVEs with shared weakness or product evidence
- CVE-2019-6540CVSS 6.5 · Medium
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Clari…
- CVE-2026-61267CVSS 7.3 · High
Vulnerability in the Oracle HCM Configuration Workbench product of Oracle E-Business Suite (component: Spreadsheet Loading). Supported versions that are affected are 12.2.3-12.2.…
- CVE-2026-42569CVSS 9.4 · Critical
phpVMS is a PHP application to run and simulate an airline. Prior to version 7.0.6, a critical vulnerability in phpVMS allowed unauthenticated access to a legacy import feature. T…
- CVE-2026-27471CVSS 9.3 · Critical
ERP is a free and open source Enterprise Resource Planning tool. In versions up to 15.98.0 and 16.0.0-rc.1 and through 16.6.0, certain endpoints lacked access validation which all…
- CVE-2022-1521CVSS 9.1 · Critical
LRM does not implement authentication or authorization by default. A malicious actor can inject, replay, modify, and/or intercept sensitive data.
- CVE-2026-73296CVSS 9.4 · Critical
Microsoft UFO open-source framework for intelligent automation across devices and platforms. Prior to 3.0.8, create_mobile_data_collection_server and create_mobile_action_server i…